How to Choose?
   
Comparisons
   
Which Voltage Sensitive Dye?
   
Minimizing Noise
   
References
 


Ask the Expert
   
Downloads
   

 
Distributors
Contact Us
Site Index








How to choose an optical imaging system for recording fast brain activity.

Optical imaging technologies are becoming more widely used in the fields of Neuroscience and Physiology. To choose an appropriate system for one's application the following aspects should be considered:


Dynamic Range and Signal-to-Noise Ratio

An important figure of merit for an optical recording system is dynamic range. Dynamic range can be specified in db, in bits, or as an exponent (e.g. 100 db = 17 bits = 105 ). The dynamic range determines the size of smallest fractional intensity change that can be measured. For example, a dynamic range of 100 db would allow one to measure a signal with a fractional change, I/I, of 10-5. The smallest signal that can be measured with a dynamic range of 60 db is 10-3. Quite often, optical signals of biological interest are small (10-2 - 10-4). A large dynamic range ensures that the apparatus can measure small signals with an optimal signal-to-noise ratio.

The signal-to-noise ratio is linearly related to the light intensity when dark noise dominates (low light) and related to the square root of the light intensity when shot noise dominates (higher light). At low light levels, as in fluorescence measurements from a single neuron or cardiac cell
, the performance of a cooled CCD camera is better than a photodiode array (low spatial resolution) or a CMOS camera (high spatial resolution). But unlike a CCD camera, the diodes or a CMOS camera won't saturate at high light levels, as in fluorescence or absorption measurements from an intact heart, vertebrate brain, brain slice, or a ganglion. As a result, CMOS cameras and photodiode systems with parallel amplification generally have a larger dynamic range, thereby yielding a better signal-to-noise ratio at high light levels. While the dynamic range cannot be larger than the effective resolution of the analogue-to-digital converter, it can be considerably smaller if, for example, saturation limits the number of measured photons. Dynamic range can also be reduced by extraneous noise.


The following is a listing of the dynamic range required for several voltage-sensitive-dye applications:
· Population signals from intact brains and brain slices -- 12-15 bits.
· Multiple individual neurons in invertebrates -- 15-17 bits.
· Multiple sites on a single cell -- 12 bits.
· Tissue cultured neurons -- 12-17 bits.
· Intact hearts -- 10 bits.


Dark Noise

The dark noise (the noise of the system with no light) affects the signal-to-noise ratio at low light levels. The dark noise is lowest in a cooled CCD. CMOS cameras and Photodiode arrays have a larger dark noise.


Temporal Resolution (Frame Rate)

A frame rate of at least 1 kHz is required for recording action potentials and other fast neuronal or cardiac signals.


Spatial Resolution

Good spatial resolution is important for obtaining a high quality image of activity. In general, increasing the number of pixels will increase the spatial resolution. However, in many preparations, light scattering or signals from out-of-focus light are substantial. Thus, a large number of pixels may not actually increase the spatial resolution, but rather will only reduce the amount of photons each pixel receives, thereby reducing the signal-to-noise ratio.


Software

The software should be comprehensive and easy to use.
For additional reading, see:
- Wu, J-Y. and Cohen, L.B. (1993). Fast multisite optical measurement of membrane potential. In: Fluorescent and Luminescent Probes for Biological Activity, W.T. Mason eds, Academic Press, London. pp 389-404.
- Grinvald, A., Frostig, R.D., Lieke E. and Hildesheim, R. (1988). Optical imaging of neuronal activity. Physiological Reviews, 68(4): 1285-1365.


Comparing technologies for improving S/N of CCD cameras:

Two different approaches could be taken to improve the signal-to-noise-ratio of a measurement at extremely low light levels, where the read noise of the chip is usually higher than the shot noise: One could increase the signal (EMCCDs), or alternately reduce the noise (MIT/LL CCD).

EMCCDs use an electron bombardment technique to amplify the signal above the readnoise floor of the CCD. The penalties for that are:

1. Spurious noise events - approximately one corrupt pixel in 1000, randomly in space and time is the very best that has been done.

2. Sqrt(2) penalty in shot noise. i.e. shot noise is sqrt(2N) instead of sqrt(N).

3. Variable gain - gain can vary pixel to pixel, line to line, frame to frame. Impossible to calibrate in an absolute sense, even in the same frame.

4. Electron bombardment process is destructive and the CCD degrades over time.

On the other hand, the low read noise MIT/LL CCD does not have any of these issues. The chip is using an ultra-low readnoise output while boosting the responsivity to 40 uV/e-. This process is simple, reliable, linear, non-destructive and constant.


Comparing signals and images from a 2-photon microscope and an ordinary fluorescence microscope (equipped with a CardioCCD-SMQ camera)

A. Calcium Green-1 signals measured with a 2-photon microscope and an ordinary fluorescence microscope with a CardioCCD-SMQ camera




B. Images


Olfactory receptor neuron nerve terminals in the mouse olfactory bulb were stained with Calcium Green-1.

Figure A. at top shows odorant elicited signals in in vivo preparations using the two imaging systems; in both cases the signals were the spatial average of the light from one glomerulus. The signal-to-noise ratio for the CardioCCD-SMQ recording is much larger. This results from a larger number of measured photons in the CardioCCD-SMQ recording. Similar results were obtained in comparisons made on five mice.

Moreover, four trials were averaged in the 2-photon measurement shown in the figure while the result from CardioCCD was from a single trial. In addition, the numerical aperture (NA) of the lens used for the 2-photon measurement was 0.8 while that used in the ordinary microscope was only 0.5. If a correction for these two factors is applied, the 2-photon measurement would have a signal-to-noise ratio six times smaller than that shown.


Factors that contribute to the relatively small number of photons in the 2-photon measurement are:

1. The incident light in the 2-photon microscope interacts with many fewer dye molecules because only a thin section receives high intensity illumination.
2. Calcium Green-1 has a 2-photon cross section which results in a low optical efficiency. This low efficiency can not be overcome by increasing the incident intensity because higher intensity will heat the preparation.

The images formed by the two kinds of microscope are shown in the bottom figure B. (the image made with the ordinary microscope covers a 2x larger area of the bulb). The 2-photon image is the total intensity; the ordinary microscope image is the image of the signal. The advantages of 2-photon microscopy are clear; rejection of scattered light and very shallow depth of focus results in much better x-y and z-axis resolution. Clearly the two kinds of imaging systems are optimal for different niches in the parameter space of imaging.

The olfactory receptor neuron staining procedure in the mouse (Wachowiak and Cohen, 2001) followed one developed by Friedrich and Korsching (1997) for zebrafish. (Data provided by Rainer Friedrich, Matt Wachowiak and Larry Cohen, Max Planck Institute for Medical Research, Heidelberg, and Yale University, New Haven.)


back to top