Cables The
cables between the CMOS camera and the control module should NEVER be
plugged or unplugged while the power is on. Unexpected Results Occasionally the camera will get in a strange mode and
the images and values will be unexpected. (This probably results from
an IDL memory leak.) First, try the "Reset the Camera" widget on the
right column of the Acquire window. Next, try exiting IDL and
restarting NeuroPlex. If that doesn't
solve the problem, try rebooting the PC and cycling the camera power.
If it is convenient you might start each recording session with a PC
reboot. Restarting Neuroplex is
recommended whenever the program behaves in an unexpected manner. Dual/Multi-core
processors The
computers supplied by us have dual/multi-core processors. In
general it is not a problem, but in certain cases (e.g. the very fast
sequences), enabling the dual/multi-core option might cause an
intermittent failure of data acquisition. If you encounter such problem, we suggest you
disable the multi-core function: Hit
the F2 key during the boot up loop and choose <Performance>
to disable the "Dual core"/"Multiple core". Interference from
other programs/Network Activity on the Computer
Attention:
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||